Abstract
Introduction: This study evaluated the QuickSee autorefractor for measurement of the accommodative response in a cohort of children without significant ametropia, and compared the findings to those determined using the gold‐standard Shin‐Nippon device. Methods: Children aged 5–7 years were recruited. QuickSee and Shin‐Nippon autorefractors were used to measure refractive status at distance (4 m) and near (50, 33 and 25 cm). Accommodative response was calculated as the difference between distance autorefraction and the value obtained for each accommodative demand. Individual accommodative response slopes were calculated from linear regression for response against demand and averaged to calculate mean slopes for each autorefractor. Inter‐instrument agreement between the QuickSee and Shin‐Nippon was assessed using the Bland–Altman method and 95% limits of agreement. Differences in measured accommodative responses between devices across accommodative demand were assessed by repeated measures ANOVA. Results: Forty‐nine children with a median (IQR) spherical equivalent refraction of +0.38 (0.50 D) (range −0.63 to +1.50 D) participated. Median (IQR) accommodative responses for the 2, 3 and 4 D demands were 1.63 D (0.69), 2.50 D (0.75) and 3.13 D (0.63), respectively, for QuickSee, and 1.63 D (0.69), 2.50 D (0.56) and 3.25 D (0.63) for the Shin‐Nippon. Mean ± SD slopes for accommodative response were 0.75 ± 0.22 and 0.83 ± 0.18 for QuickSee and Shin‐Nippon, respectively (p = 0.09). Pearson's correlation showed no significant proportional bias for measures of accommodative response with either autorefractor (p ≥ 0.19). The mean difference in accommodative response measured by the QuickSee and Shin‐Nippon did not differ significantly with accommodative demand (F = 1.70, p = 0.19). Conclusion: The agreement observed here demonstrates that the QuickSee shows promise in measuring accommodative response over a range of demands (2, 3 and 4 D) in a paediatric population without significant ametropia. Further research is required to determine if these findings translate over a wider range of refractive errors, and to evaluate the value of QuickSee accommodative response measures in vision screening.
| Original language | English |
|---|---|
| Pages (from-to) | 1581-1590 |
| Number of pages | 10 |
| Journal | Ophthalmic and Physiological Optics |
| Volume | 45 |
| Issue number | 7 |
| Early online date | 19 Jul 2025 |
| DOIs | |
| Publication status | Published (in print/issue) - 30 Nov 2025 |
Bibliographical note
Publisher Copyright:© 2025 The Author(s). Ophthalmic and Physiological Optics published by John Wiley & Sons Ltd on behalf of College of Optometrists.
Funding
The study was performed with the in‐kind support of equipment and resources from the Centre for Optometry and Vision Science (School of Biomedical Sciences, Ulster University) for the research associate and team; no additional funding was provided.
UN SDGs
This output contributes to the following UN Sustainable Development Goals (SDGs)
-
SDG 3 Good Health and Well-being
Keywords
- vision screening
- refractive error
- hyperopia
- children
- accommodation
- QuickSee
- Reproducibility of Results
- Humans
- Child, Preschool
- Male
- Refractive Errors/physiopathology
- Equipment Design
- Accommodation, Ocular/physiology
- Female
- Refraction, Ocular/physiology
- Child
- Vision Tests/instrumentation
Fingerprint
Dive into the research topics of 'The QuickSee autorefractor reliably measures the accommodative response in children compared with the Shin‐Nippon autorefractor'. Together they form a unique fingerprint.Student theses
-
An evaluation of the prevalence of refractive error and the relationship between relative peripheral refraction and eye growth in children
Leighton, R. E. (Author), Saunders, K. (Supervisor), Mc Cullough, S. (Supervisor) & Breslin, K. (Supervisor), Jun 2022Student thesis: Doctoral Thesis
File -
Towards seeing clearly: the contribution of atypical optical structures to refractive error and accommodative function in Down syndrome
Uwacu, T. D. F. (Author), Bhattacharyya, S. (Supervisor), Saunders, K. (Supervisor), Little, J.-A. (Supervisor) & Gonzalez-Meijome, J. M. (Supervisor), Jun 2026Student thesis: Doctoral Thesis
File