Piezoelectric sensor determination of arterial pulse wave velocity

JAD McLaughlin, MDJ McNeill, B Braun, PD McCormack

Research output: Contribution to journalArticle

51 Citations (Scopus)

Abstract

Arterial pulse wave velocity (APWV) is a measure of the elasticity (or stiffness) of peripheral arterial blood vessels. The pulse referred to here will be the pressure pulse as opposed to the flow pulse measured by ultrasound Doppler. The pressure pulse velocity varies over the range from about 12 m s−1 to 15 m s−1 in stiff peripheral arteries, whereas in normal arteries it has a velocity in the range of 7 to 9 m s−1.The aim of this project was the development of a fast and easy to use system for the determination of peripheral arterial pulse wave velocity. The principle of the PWV measurement is based on simultaneous measurement of two pulse waves at two different positions, such as the radial artery at the wrist and the brachial artery just above the elbow. By determining the pulse transit time between these points and the distance measured between the two locations, pulse wave velocity may then be calculated. The pressure pulse detection is done by using two piezoelectric sensors which generate a measurable voltage at the output contacts if they are mechanically deformed. The deformation produced voltage is first amplified and filtered and then digitalized with a data acquisition card. The analysis of the data obtained from the sensors includes a filtering process, the calculation of the PWV with three different methods—foot-to-foot, cross-correlation and peak-to-peak—and the determination of the arterial pulse rate.Extensive measurements with human test subjects were carried out to optimize the techniques of data acquisition and analysis. For example, it was found that the best procedure was to hold the sensors in place using elastic straps alone. The data analysis was upgraded with an additional software module, which deletes, in effect, outriders or invalid measurements. With the optimized system, a series involving eight test subjects ranging in age from 22 to 32 years was completed (all normotensive). The arterial pulse wave velocities determined covered a range from 6 m s−1 to 12 m s−1, with an average standard deviation of less than 2.5 m s−1 for individual results. These are slightly higher, but close to published APWV data. The results showed that reproducible results can be obtained with the existing PWV acquirement and analysis system.Copyright IOP 2003
LanguageEnglish
Pages693-702
JournalPhysiological Measurement
Volume24
Issue number3
DOIs
Publication statusPublished - 2003

Fingerprint

Pulse Wave Analysis
Sensors
Blood Pressure
Foot
Arteries
Data acquisition
Doppler Ultrasonography
Vascular Stiffness
Radial Artery
Brachial Artery
Elasticity
Elbow
Wrist
Blood vessels
Electric potential
Blood Vessels
Pulse
Software
Ultrasonics
Stiffness

Keywords

  • arterial pulse wave velocity
  • APWV
  • non-invasive
  • pressure pulse
  • piezoelectric sensor
  • elasticity
  • stiffness
  • peripheral

Cite this

McLaughlin, JAD ; McNeill, MDJ ; Braun, B ; McCormack, PD. / Piezoelectric sensor determination of arterial pulse wave velocity. In: Physiological Measurement. 2003 ; Vol. 24, No. 3. pp. 693-702.
@article{a66a8ea569f7461eb33bf886916d7e09,
title = "Piezoelectric sensor determination of arterial pulse wave velocity",
abstract = "Arterial pulse wave velocity (APWV) is a measure of the elasticity (or stiffness) of peripheral arterial blood vessels. The pulse referred to here will be the pressure pulse as opposed to the flow pulse measured by ultrasound Doppler. The pressure pulse velocity varies over the range from about 12 m s−1 to 15 m s−1 in stiff peripheral arteries, whereas in normal arteries it has a velocity in the range of 7 to 9 m s−1.The aim of this project was the development of a fast and easy to use system for the determination of peripheral arterial pulse wave velocity. The principle of the PWV measurement is based on simultaneous measurement of two pulse waves at two different positions, such as the radial artery at the wrist and the brachial artery just above the elbow. By determining the pulse transit time between these points and the distance measured between the two locations, pulse wave velocity may then be calculated. The pressure pulse detection is done by using two piezoelectric sensors which generate a measurable voltage at the output contacts if they are mechanically deformed. The deformation produced voltage is first amplified and filtered and then digitalized with a data acquisition card. The analysis of the data obtained from the sensors includes a filtering process, the calculation of the PWV with three different methods—foot-to-foot, cross-correlation and peak-to-peak—and the determination of the arterial pulse rate.Extensive measurements with human test subjects were carried out to optimize the techniques of data acquisition and analysis. For example, it was found that the best procedure was to hold the sensors in place using elastic straps alone. The data analysis was upgraded with an additional software module, which deletes, in effect, outriders or invalid measurements. With the optimized system, a series involving eight test subjects ranging in age from 22 to 32 years was completed (all normotensive). The arterial pulse wave velocities determined covered a range from 6 m s−1 to 12 m s−1, with an average standard deviation of less than 2.5 m s−1 for individual results. These are slightly higher, but close to published APWV data. The results showed that reproducible results can be obtained with the existing PWV acquirement and analysis system.Copyright IOP 2003",
keywords = "arterial pulse wave velocity, APWV, non-invasive, pressure pulse, piezoelectric sensor, elasticity, stiffness, peripheral",
author = "JAD McLaughlin and MDJ McNeill and B Braun and PD McCormack",
year = "2003",
doi = "10.1088/0967-3334/24/3/306",
language = "English",
volume = "24",
pages = "693--702",
journal = "Physiological Measurement",
issn = "0967-3334",
number = "3",

}

Piezoelectric sensor determination of arterial pulse wave velocity. / McLaughlin, JAD; McNeill, MDJ; Braun, B; McCormack, PD.

In: Physiological Measurement, Vol. 24, No. 3, 2003, p. 693-702.

Research output: Contribution to journalArticle

TY - JOUR

T1 - Piezoelectric sensor determination of arterial pulse wave velocity

AU - McLaughlin, JAD

AU - McNeill, MDJ

AU - Braun, B

AU - McCormack, PD

PY - 2003

Y1 - 2003

N2 - Arterial pulse wave velocity (APWV) is a measure of the elasticity (or stiffness) of peripheral arterial blood vessels. The pulse referred to here will be the pressure pulse as opposed to the flow pulse measured by ultrasound Doppler. The pressure pulse velocity varies over the range from about 12 m s−1 to 15 m s−1 in stiff peripheral arteries, whereas in normal arteries it has a velocity in the range of 7 to 9 m s−1.The aim of this project was the development of a fast and easy to use system for the determination of peripheral arterial pulse wave velocity. The principle of the PWV measurement is based on simultaneous measurement of two pulse waves at two different positions, such as the radial artery at the wrist and the brachial artery just above the elbow. By determining the pulse transit time between these points and the distance measured between the two locations, pulse wave velocity may then be calculated. The pressure pulse detection is done by using two piezoelectric sensors which generate a measurable voltage at the output contacts if they are mechanically deformed. The deformation produced voltage is first amplified and filtered and then digitalized with a data acquisition card. The analysis of the data obtained from the sensors includes a filtering process, the calculation of the PWV with three different methods—foot-to-foot, cross-correlation and peak-to-peak—and the determination of the arterial pulse rate.Extensive measurements with human test subjects were carried out to optimize the techniques of data acquisition and analysis. For example, it was found that the best procedure was to hold the sensors in place using elastic straps alone. The data analysis was upgraded with an additional software module, which deletes, in effect, outriders or invalid measurements. With the optimized system, a series involving eight test subjects ranging in age from 22 to 32 years was completed (all normotensive). The arterial pulse wave velocities determined covered a range from 6 m s−1 to 12 m s−1, with an average standard deviation of less than 2.5 m s−1 for individual results. These are slightly higher, but close to published APWV data. The results showed that reproducible results can be obtained with the existing PWV acquirement and analysis system.Copyright IOP 2003

AB - Arterial pulse wave velocity (APWV) is a measure of the elasticity (or stiffness) of peripheral arterial blood vessels. The pulse referred to here will be the pressure pulse as opposed to the flow pulse measured by ultrasound Doppler. The pressure pulse velocity varies over the range from about 12 m s−1 to 15 m s−1 in stiff peripheral arteries, whereas in normal arteries it has a velocity in the range of 7 to 9 m s−1.The aim of this project was the development of a fast and easy to use system for the determination of peripheral arterial pulse wave velocity. The principle of the PWV measurement is based on simultaneous measurement of two pulse waves at two different positions, such as the radial artery at the wrist and the brachial artery just above the elbow. By determining the pulse transit time between these points and the distance measured between the two locations, pulse wave velocity may then be calculated. The pressure pulse detection is done by using two piezoelectric sensors which generate a measurable voltage at the output contacts if they are mechanically deformed. The deformation produced voltage is first amplified and filtered and then digitalized with a data acquisition card. The analysis of the data obtained from the sensors includes a filtering process, the calculation of the PWV with three different methods—foot-to-foot, cross-correlation and peak-to-peak—and the determination of the arterial pulse rate.Extensive measurements with human test subjects were carried out to optimize the techniques of data acquisition and analysis. For example, it was found that the best procedure was to hold the sensors in place using elastic straps alone. The data analysis was upgraded with an additional software module, which deletes, in effect, outriders or invalid measurements. With the optimized system, a series involving eight test subjects ranging in age from 22 to 32 years was completed (all normotensive). The arterial pulse wave velocities determined covered a range from 6 m s−1 to 12 m s−1, with an average standard deviation of less than 2.5 m s−1 for individual results. These are slightly higher, but close to published APWV data. The results showed that reproducible results can be obtained with the existing PWV acquirement and analysis system.Copyright IOP 2003

KW - arterial pulse wave velocity

KW - APWV

KW - non-invasive

KW - pressure pulse

KW - piezoelectric sensor

KW - elasticity

KW - stiffness

KW - peripheral

U2 - 10.1088/0967-3334/24/3/306

DO - 10.1088/0967-3334/24/3/306

M3 - Article

VL - 24

SP - 693

EP - 702

JO - Physiological Measurement

T2 - Physiological Measurement

JF - Physiological Measurement

SN - 0967-3334

IS - 3

ER -