Macrophage responses to vascular stent coatings

M Ball, A O'Brien, F Dolan, GA Abbas, JAD McLaughlin

Research output: Contribution to journalArticle

35 Citations (Scopus)

Abstract

Diamond-like carbon (DLC) films have been proposed as potential coatings for blood-contacting devices. In this study, tetrahedral amorphous carbon (ta-C) films deposited by filtered cathodic vacuum arc system (FCVA) were compared with commercially deposited polyurethane coatings (PU) and uncoated stainless steel samples. X-ray reflectivity (XRR) measurements were performed to check density and thickness of the ta-C coatings, and contact angles measurements were used to assess surface wettability. J774 macrophages were used to assess the cell responses to the materials. Cell number, metabolic activity, and hydrogen peroxide production were measured by using biochemical assays, and the cell attachment and morphology were determined by using scanning electron microscopy and confocal laser scanning microscopy. Wettability measurements showed that of the materials, the stainless steel was the most hydrophilic, and the ta-C coatings were the most hydrophobic. Although the initial attachment and morphology did notappear to be dependent on the wettability, the cell numbers did increase with increasing wettability. Macrophages on the stainless steel samples were the most active in producing hydrogen peroxide. These data show that ta-C samples performed as well as commercialPU-coated samples in blocking cell reactions to the substrate and may prove to be effective coatings for blood-contacting materials. (C) 2004 Wiley Periodicals, Inc.
LanguageEnglish
Pages380-390
JournalJournal of Biomedical Materials Research Part a
Volume70A
Issue number3
DOIs
Publication statusPublished - Sep 2004

Fingerprint

Stents
Macrophages
Amorphous carbon
Wetting
Stainless Steel
Coatings
Hydrogen Peroxide
Blood
Diamond like carbon films
Polyurethanes
Carbon films
Amorphous films
Angle measurement
Contact angle
Assays
Microscopic examination
Vacuum
Scanning
X rays
Scanning electron microscopy

Keywords

  • thin film coating
  • tetrahedral amorphous carbon
  • cell adhesion
  • inflammation
  • macrophages

Cite this

Ball, M ; O'Brien, A ; Dolan, F ; Abbas, GA ; McLaughlin, JAD. / Macrophage responses to vascular stent coatings. In: Journal of Biomedical Materials Research Part a. 2004 ; Vol. 70A, No. 3. pp. 380-390.
@article{918de8f4c71b4c80b6976e2384c0067b,
title = "Macrophage responses to vascular stent coatings",
abstract = "Diamond-like carbon (DLC) films have been proposed as potential coatings for blood-contacting devices. In this study, tetrahedral amorphous carbon (ta-C) films deposited by filtered cathodic vacuum arc system (FCVA) were compared with commercially deposited polyurethane coatings (PU) and uncoated stainless steel samples. X-ray reflectivity (XRR) measurements were performed to check density and thickness of the ta-C coatings, and contact angles measurements were used to assess surface wettability. J774 macrophages were used to assess the cell responses to the materials. Cell number, metabolic activity, and hydrogen peroxide production were measured by using biochemical assays, and the cell attachment and morphology were determined by using scanning electron microscopy and confocal laser scanning microscopy. Wettability measurements showed that of the materials, the stainless steel was the most hydrophilic, and the ta-C coatings were the most hydrophobic. Although the initial attachment and morphology did notappear to be dependent on the wettability, the cell numbers did increase with increasing wettability. Macrophages on the stainless steel samples were the most active in producing hydrogen peroxide. These data show that ta-C samples performed as well as commercialPU-coated samples in blocking cell reactions to the substrate and may prove to be effective coatings for blood-contacting materials. (C) 2004 Wiley Periodicals, Inc.",
keywords = "thin film coating, tetrahedral amorphous carbon, cell adhesion, inflammation, macrophages",
author = "M Ball and A O'Brien and F Dolan and GA Abbas and JAD McLaughlin",
year = "2004",
month = "9",
doi = "10.1002/jbm.a.30066",
language = "English",
volume = "70A",
pages = "380--390",
journal = "Journal of Biomedical Materials Research: Part A",
issn = "1549-3296",
number = "3",

}

Macrophage responses to vascular stent coatings. / Ball, M; O'Brien, A; Dolan, F; Abbas, GA; McLaughlin, JAD.

In: Journal of Biomedical Materials Research Part a, Vol. 70A, No. 3, 09.2004, p. 380-390.

Research output: Contribution to journalArticle

TY - JOUR

T1 - Macrophage responses to vascular stent coatings

AU - Ball, M

AU - O'Brien, A

AU - Dolan, F

AU - Abbas, GA

AU - McLaughlin, JAD

PY - 2004/9

Y1 - 2004/9

N2 - Diamond-like carbon (DLC) films have been proposed as potential coatings for blood-contacting devices. In this study, tetrahedral amorphous carbon (ta-C) films deposited by filtered cathodic vacuum arc system (FCVA) were compared with commercially deposited polyurethane coatings (PU) and uncoated stainless steel samples. X-ray reflectivity (XRR) measurements were performed to check density and thickness of the ta-C coatings, and contact angles measurements were used to assess surface wettability. J774 macrophages were used to assess the cell responses to the materials. Cell number, metabolic activity, and hydrogen peroxide production were measured by using biochemical assays, and the cell attachment and morphology were determined by using scanning electron microscopy and confocal laser scanning microscopy. Wettability measurements showed that of the materials, the stainless steel was the most hydrophilic, and the ta-C coatings were the most hydrophobic. Although the initial attachment and morphology did notappear to be dependent on the wettability, the cell numbers did increase with increasing wettability. Macrophages on the stainless steel samples were the most active in producing hydrogen peroxide. These data show that ta-C samples performed as well as commercialPU-coated samples in blocking cell reactions to the substrate and may prove to be effective coatings for blood-contacting materials. (C) 2004 Wiley Periodicals, Inc.

AB - Diamond-like carbon (DLC) films have been proposed as potential coatings for blood-contacting devices. In this study, tetrahedral amorphous carbon (ta-C) films deposited by filtered cathodic vacuum arc system (FCVA) were compared with commercially deposited polyurethane coatings (PU) and uncoated stainless steel samples. X-ray reflectivity (XRR) measurements were performed to check density and thickness of the ta-C coatings, and contact angles measurements were used to assess surface wettability. J774 macrophages were used to assess the cell responses to the materials. Cell number, metabolic activity, and hydrogen peroxide production were measured by using biochemical assays, and the cell attachment and morphology were determined by using scanning electron microscopy and confocal laser scanning microscopy. Wettability measurements showed that of the materials, the stainless steel was the most hydrophilic, and the ta-C coatings were the most hydrophobic. Although the initial attachment and morphology did notappear to be dependent on the wettability, the cell numbers did increase with increasing wettability. Macrophages on the stainless steel samples were the most active in producing hydrogen peroxide. These data show that ta-C samples performed as well as commercialPU-coated samples in blocking cell reactions to the substrate and may prove to be effective coatings for blood-contacting materials. (C) 2004 Wiley Periodicals, Inc.

KW - thin film coating

KW - tetrahedral amorphous carbon

KW - cell adhesion

KW - inflammation

KW - macrophages

U2 - 10.1002/jbm.a.30066

DO - 10.1002/jbm.a.30066

M3 - Article

VL - 70A

SP - 380

EP - 390

JO - Journal of Biomedical Materials Research: Part A

T2 - Journal of Biomedical Materials Research: Part A

JF - Journal of Biomedical Materials Research: Part A

SN - 1549-3296

IS - 3

ER -