Abstract
A new device has been developed for the application of Electrical Stimulation to improve healing of chronic wounds. The device enables the creation of a composite electrode hence matching the electrode(s) to the size and shape of the wound. Up to 49 electrodes in an array can be combined, delivering High Voltage Pulsed Current (HVPC) in the range 60-120 Hz with a pulse duration range from 90-110 mus and the possibility of treating the patient with direct current instead of HVPC. In addition, the software can import the measurements from the ImpediMap device, analysing the electrical impedance of the tissues involved. A test on healthy volunteers did not prove a statistically significant rise in skin temperature, TcPO2 or impedance due to the stimulation, even though a slight reddening of the stimulated site was observed.
Original language | English |
---|---|
Title of host publication | Unknown Host Publication |
Publisher | IEEE Engineering in Medicine and Biology Society |
Pages | 2145-8 |
Number of pages | 4 |
Volume | 2009 |
Publication status | Published (in print/issue) - 2009 |
Event | Annual International Conference of the IEEE Engineering in Medicine and Biology Society - Minneapolis Duration: 1 Jan 2009 → … |
Conference
Conference | Annual International Conference of the IEEE Engineering in Medicine and Biology Society |
---|---|
Period | 1/01/09 → … |