Detection of Cardiac Arrest using a Simplified Frequency Analysis of the Impedance Cardiogram recorded from Defibrillator Pads

CO Navarro, NA Cromie, C Turner, JMCC Anderson

Research output: Chapter in Book/Report/Conference proceedingConference contribution

1 Citation (Scopus)

Abstract

An algorithm based only on the impedance cardiogram (ICG) recorded through two defibrillation pads, using the strongest frequency component and amplitude, incorporated into a defibrillator could determine circulatory arrest and reduce delays in starting cardiopulmonary resuscitation (CPR). Frequency analysis of the ICG signal is carried out by integer filters on a sample by sample basis. They are simpler, lighter and more versatile when compared to the FFT. This alternative approach, although less accurate, is preferred due to the limited processing capacity of devices that could compromise real time usability of the FFT. These two techniques were compared across a data set comprising 13 cases of cardiac arrest and 6 normal controls. The best filters were refined on this training set and an algorithm for the detection of cardiac arrest was trained on a wider data set. The algorithm was finally tested on a validation set. The ICG was recorded in 132 cardiac arrest patients (53 training, 79 validation) and 97 controls (47 training, 50 validation): the diagnostic algorithm indicated cardiac arrest with a sensitivity of 81.1% (77.6 – 84.3) and specificity of 97.1% (96.7 – 97.4) for the validation set (95% confidence intervals). Automated defibrillators with integrated ICG analysis have the potential to improve emergency care by lay persons enabling more rapid and appropriate initiation of CPR and when combined with ECG analysis they could improve on the detection of cardiac arrest.
LanguageEnglish
Title of host publicationUnknown Host Publication
Pages1709-1712
Number of pages4
DOIs
Publication statusPublished - 29 Dec 2011
Event33rd Annual International Conference of the IEEE EMBS - Boston, MA, USA
Duration: 29 Dec 2011 → …

Conference

Conference33rd Annual International Conference of the IEEE EMBS
Period29/12/11 → …

Fingerprint

Defibrillators
Heart Arrest
Electric Impedance
Cardiopulmonary Resuscitation
Emergency Medical Services
Electrocardiography
Confidence Intervals
Equipment and Supplies

Keywords

  • ICG
  • bioimpedance
  • resuscitation
  • impedance cardiogram
  • CPR
  • defibrillation
  • FFT
  • digital filtering

Cite this

@inproceedings{f164749aff8c4b969cf066e98ae58120,
title = "Detection of Cardiac Arrest using a Simplified Frequency Analysis of the Impedance Cardiogram recorded from Defibrillator Pads",
abstract = "An algorithm based only on the impedance cardiogram (ICG) recorded through two defibrillation pads, using the strongest frequency component and amplitude, incorporated into a defibrillator could determine circulatory arrest and reduce delays in starting cardiopulmonary resuscitation (CPR). Frequency analysis of the ICG signal is carried out by integer filters on a sample by sample basis. They are simpler, lighter and more versatile when compared to the FFT. This alternative approach, although less accurate, is preferred due to the limited processing capacity of devices that could compromise real time usability of the FFT. These two techniques were compared across a data set comprising 13 cases of cardiac arrest and 6 normal controls. The best filters were refined on this training set and an algorithm for the detection of cardiac arrest was trained on a wider data set. The algorithm was finally tested on a validation set. The ICG was recorded in 132 cardiac arrest patients (53 training, 79 validation) and 97 controls (47 training, 50 validation): the diagnostic algorithm indicated cardiac arrest with a sensitivity of 81.1{\%} (77.6 – 84.3) and specificity of 97.1{\%} (96.7 – 97.4) for the validation set (95{\%} confidence intervals). Automated defibrillators with integrated ICG analysis have the potential to improve emergency care by lay persons enabling more rapid and appropriate initiation of CPR and when combined with ECG analysis they could improve on the detection of cardiac arrest.",
keywords = "ICG, bioimpedance, resuscitation, impedance cardiogram, CPR, defibrillation, FFT, digital filtering",
author = "CO Navarro and NA Cromie and C Turner and JMCC Anderson",
year = "2011",
month = "12",
day = "29",
doi = "10.1109/IEMBS.2011.6090490",
language = "English",
pages = "1709--1712",
booktitle = "Unknown Host Publication",

}

Navarro, CO, Cromie, NA, Turner, C & Anderson, JMCC 2011, Detection of Cardiac Arrest using a Simplified Frequency Analysis of the Impedance Cardiogram recorded from Defibrillator Pads. in Unknown Host Publication. pp. 1709-1712, 33rd Annual International Conference of the IEEE EMBS, 29/12/11. https://doi.org/10.1109/IEMBS.2011.6090490

Detection of Cardiac Arrest using a Simplified Frequency Analysis of the Impedance Cardiogram recorded from Defibrillator Pads. / Navarro, CO; Cromie, NA; Turner, C; Anderson, JMCC.

Unknown Host Publication. 2011. p. 1709-1712.

Research output: Chapter in Book/Report/Conference proceedingConference contribution

TY - GEN

T1 - Detection of Cardiac Arrest using a Simplified Frequency Analysis of the Impedance Cardiogram recorded from Defibrillator Pads

AU - Navarro, CO

AU - Cromie, NA

AU - Turner, C

AU - Anderson, JMCC

PY - 2011/12/29

Y1 - 2011/12/29

N2 - An algorithm based only on the impedance cardiogram (ICG) recorded through two defibrillation pads, using the strongest frequency component and amplitude, incorporated into a defibrillator could determine circulatory arrest and reduce delays in starting cardiopulmonary resuscitation (CPR). Frequency analysis of the ICG signal is carried out by integer filters on a sample by sample basis. They are simpler, lighter and more versatile when compared to the FFT. This alternative approach, although less accurate, is preferred due to the limited processing capacity of devices that could compromise real time usability of the FFT. These two techniques were compared across a data set comprising 13 cases of cardiac arrest and 6 normal controls. The best filters were refined on this training set and an algorithm for the detection of cardiac arrest was trained on a wider data set. The algorithm was finally tested on a validation set. The ICG was recorded in 132 cardiac arrest patients (53 training, 79 validation) and 97 controls (47 training, 50 validation): the diagnostic algorithm indicated cardiac arrest with a sensitivity of 81.1% (77.6 – 84.3) and specificity of 97.1% (96.7 – 97.4) for the validation set (95% confidence intervals). Automated defibrillators with integrated ICG analysis have the potential to improve emergency care by lay persons enabling more rapid and appropriate initiation of CPR and when combined with ECG analysis they could improve on the detection of cardiac arrest.

AB - An algorithm based only on the impedance cardiogram (ICG) recorded through two defibrillation pads, using the strongest frequency component and amplitude, incorporated into a defibrillator could determine circulatory arrest and reduce delays in starting cardiopulmonary resuscitation (CPR). Frequency analysis of the ICG signal is carried out by integer filters on a sample by sample basis. They are simpler, lighter and more versatile when compared to the FFT. This alternative approach, although less accurate, is preferred due to the limited processing capacity of devices that could compromise real time usability of the FFT. These two techniques were compared across a data set comprising 13 cases of cardiac arrest and 6 normal controls. The best filters were refined on this training set and an algorithm for the detection of cardiac arrest was trained on a wider data set. The algorithm was finally tested on a validation set. The ICG was recorded in 132 cardiac arrest patients (53 training, 79 validation) and 97 controls (47 training, 50 validation): the diagnostic algorithm indicated cardiac arrest with a sensitivity of 81.1% (77.6 – 84.3) and specificity of 97.1% (96.7 – 97.4) for the validation set (95% confidence intervals). Automated defibrillators with integrated ICG analysis have the potential to improve emergency care by lay persons enabling more rapid and appropriate initiation of CPR and when combined with ECG analysis they could improve on the detection of cardiac arrest.

KW - ICG

KW - bioimpedance

KW - resuscitation

KW - impedance cardiogram

KW - CPR

KW - defibrillation

KW - FFT

KW - digital filtering

U2 - 10.1109/IEMBS.2011.6090490

DO - 10.1109/IEMBS.2011.6090490

M3 - Conference contribution

SP - 1709

EP - 1712

BT - Unknown Host Publication

ER -